The MI 1005 medical network isolation module is suitable for the use in 10/100/1000-BaseT-Ethernet Gigabit networks with full IEEE 802.3 compatibility.
If PCs or other electrical devices are connected to an Ethernet network in the patient's environment, inadmissible leakage currents can appear from other devices that are also connected to network. The MI 1005 blocks these currents and voltages over the copper network. The MI 1005 device has been manufactured using UL94-V0 materials only. Compared to other network isolation devices that are available on the market, the MI 1005 is the only device available with a signal isolation strength of 5000 Volts within full IEEE 802.3 Gigabit Network support. More info.
Complies with:
- IEC / DIN EN 60601-1 3rd Edition
- IEC / DIN EN 60601-1-2:2007
Additional Documents:
MI 1005 Insulation Statement (PDF)
Product Information (PDF)
MI 1005 Manual (PDF)
Article No. | 2005674 |
Data Speed | 10/100/1000 MBit/s |
Isolation Voltage | 5000 Volts AC (50/60Hz 1 Minute) |
Connection sockets | 2 x RJ45 |
Network specifications | IEEE 802.3 10/100/1000-BaseT, Twisted-Pair, auto-conf (completely transparent in ethernet network, no drivers necessary) |
Material | UL94V-0 complying materials RoHs compliant |
Mark of Origin (Manufacturer) | Baaske Datentechnik, 32361 Preussisch Oldendorf Germany |
|
Type Designation | MI 1005 | |
Operating Mode | Continuous operation | |
Environmental conditions during operation | Temperature | -10°C bis +85°C |
Relative air humidity | 10% bis 90% (not condensing!) | |
Air pressure | 860hPa bis 1060hPa | |
Environmental conditions during Storage/Transport | Temperature | -25°C bis +85°C |
Relative air humidity | 10% bis 95% (not condensing!) | |
Air pressure | 500hPa bis 1060hPa | |
Dimenstion (LxBxH) | (65x29x23)mm³ | |
Weight | ca. 50g | |
Classification | Protection Class | Not applicable |
Protection against water penetration | No protection | |
Degree of protection of the user parts against electric shock | Not applicable | |
Degree of protection against discharge from defibrillators | Not applicable | |
Protection during use in the presence of combustible compounds of anaesthetics with air or with oxygen or laughing gas | No protection | |
Connection to the supply | Not applicable | |
Supply frequency | Not applicable | |
Power input | Not applicable | |
Pollution degree | 2 |